西門子彩超正在使用時出現關機死機維修技巧
西門子彩超開機(ji)(ji)(ji)緩慢(man)大(da)約15分鐘能開機(ji)(ji)(ji)器(qi)(在運行到Configuration時(shi)(shi)(shi)候不動(dong)(dong))維(wei)修(xiu), 進(jin)入(ru)(ru)二維(wei)圖(tu)(tu)像時(shi)(shi)(shi)死機(ji)(ji)(ji)維(wei)修(xiu),工作站圖(tu)(tu)像無法傳輸維(wei)修(xiu), 開機(ji)(ji)(ji)時(shi)(shi)(shi)進(jin)入(ru)(ru)二維(wei)圖(tu)(tu)像時(shi)(shi)(shi)好報(bao)(bao)錯(cuo)(cuo)維(wei)修(xiu):The system is shutting down automatically to prevent permanent damage,please contact Siemens Service.點擊確(que)定自動(dong)(dong)關機(ji)(ji)(ji),這種錯(cuo)(cuo)誤(wu)很多公司很難(nan)判斷,任何硬件壞(huai)都會報(bao)(bao)此(ci)錯(cuo)(cuo)誤(wu),他們越(yue)修(xiu)越(yue)麻(ma)煩,導(dao)致機(ji)(ji)(ji)器(qi)癱瘓。 機(ji)(ji)(ji)器(qi)在切換探(tan)頭報(bao)(bao)錯(cuo)(cuo)The system is shutting down automatically to prevent permanent damage,please contact Siemens Service維(wei)修(xiu),機(ji)(ji)(ji)器(qi)開機(ji)(ji)(ji)時(shi)(shi)(shi)開始運行WINDOWS XP時(shi)(shi)(shi)需要輸入(ru)(ru)用戶和密(mi)碼維(wei)修(xiu)
先懷(huai)疑HV問題,高壓是由(you)DCDC板(ban)(ban)供給的,在機(ji)器(qi)左側舒(shu)淇的板(ban)(ban)子(zi)就是DCDC板(ban)(ban),48V輸入(ru),HV是100V,用表(biao)測量100V的動態,當機(ji)器(qi)出現報錯后(hou)100V是否一直存在
西門子彩超正在使用時出現關機死機維修技巧4,如果100V停了,說明DCDC肯定有問題,如果100V還一直有,那么TI板的可能性就很大了
5,當報錯沒出現的時候,超聲區出現水波紋的橫向干擾DC板的問題比較大
6,當報錯沒出現的時候,超聲探頭的聲音很大DC板的問題比較大
7,測試機器還有一個辦法,就是讓機器進入維修界面,進去維修界面后不要退出來讓機器一直運行。如果機器有問題,一般不會出現對話框,機器也不會關機,當內部出問題的時候,退出維修界面,超聲區已經沒有回聲了,這個時候就可以用表量各個點的參數,跟正常的機器狀態進行比較。
8,DC板維修,可以進行脫機維修,接上48V電源,量各個地方的參數是否正常
9,TI板脫機維修,TI板引起的故障原因相當多,5,6處地方都會有可能性,BGA問題的話就麻煩一些,這個涉及到技術問題我就不詳細闡述,可以維修技術
10,3個(ge)探頭都(dou)出現放射狀的(de)干擾(rao),仿佛在(zai)(zai)圖像(xiang)(xiang)上蒙上一層(ceng)靜態(tai)的(de)干擾(rao),干擾(rao)層(ceng)跟圖像(xiang)(xiang)是分開的(de),增(zeng)益調試(shi)不影(ying)響圖像(xiang)(xiang)的(de)變化。這個(ge)故障(zhang)大多數是TI板接觸(chu)不好,重(zhong)新安裝下可能就好了。聲。在(zai)(zai)海外設立或者入(ru)股(gu)了很多分公司(si)。